All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Abstract

Irreversible Biamperometry Determination Of Penicillamine Based On A Flow-Injection Analysis

Author(s): Li Li-Jun, Cheng Hao, Chen Qi-Feng, Huang Wen-Yi

An original flow-injection coupling irreversible biamperometric method is developed for the direct determination of penicillamine. The method depends on electrocatalytic oxidation of penicillamine at a gold electrode and the reduction of permanganate at a platinum electrode to form an irreversible biamperometric detection system. Under the applied potential difference of 0 V, in the 0.05mol/L sulfuric acid, penicillamine can be determined over the range 6.00×10-5-1.00×10-3 mol L-1with a sampling frequency of 80 samples per hour. The detection limit for penicillamine is 8.0×10-6 mol L-1 and the RSD for 19 replicate determinations of 4.0× 10-4 mol/L penicillamine is 2.89%. The proposed method was applied to the analysis of penicillamine in penicillamine tablets with satisfactory results.