All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Abstract

Development of RP-HPLC Method for Determination of Ezetimibe and Fenofibrate from Synthetic Mixture

Author(s): S. Sambath Kumar, Suman Jain, S. Kannan, Rahul Dixit, Bhabatosh Halder and V. K. Rai

A simple precise, accurate and validated reverse phase HPLC method has been developed for the simultaneous estimation of ezetimibe and fenofibrate in bulk and in synthetic mixture using C18 column (Kromosil, 4.6 mm x 25 cm, 5 μm) with acetonitrile: 0.05 M ammonium acetate buffer (85 : 15 v/v) as a mobile phase, at a flow rate of 1.3 mL/min and detection was done at 253.0 nm. The retention time for ezetimibe and fenofibrate was found to be 2.41 ± 0.011 and 6.03 ± 0.023 min, respectively. Linearity of ezetimibe and fenofibrate was found in the range of 2-20 μg/mL and 16-80 μg/mL, respectively. The percentage assay of ezetimibe and fenofibrate was found between 99% to 101%. The statistical parameters were found within range.


Share this       
Google Scholar citation report
Citations : 9398

International Journal of Chemical Sciences received 9398 citations as per Google Scholar report

Indexed In

  • Google Scholar
  • Open J Gate
  • China National Knowledge Infrastructure (CNKI)
  • Cosmos IF
  • Geneva Foundation for Medical Education and Research
  • ICMJE

View More