All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Abstract

Determination of 3-Trifluoromethyl benzaldehydegenotoxic impurity in Cinnacalcet drug substances

Author(s): S.Senthil kumar, Sushil jaiswal,V.Srinivasarao

A new simple, sensitive HPLC method developed and validated for the determination of 3- Triflouromethyl benzaldehyde in Cinacalcet. 3- Triflouromethyl benzaldehyde was separated on Inertcil ODS-3V analytical column with water and acetonitrile used as mobile phase at a flow rate 1.0ml/min. The effluent was monitored by UV detection at 240nm. Column temperature maintained at 35ºC, injection volume 25µl. Calibration plots were linear in the range of LOQ to 150% level and LOQ and LOD were 0.06 and 0.02.The high recovery and low relative standard deviation confirm the suitability of the method for routine quality control determination of 3- Triflouromethyl benzaldehyde in Cinnacalcet